Siemens SITRANS FC430 Operating Instructions Manual page 454

Coriolis flowmeters
Hide thumbs Also See for SITRANS FC430:
Table of Contents

Advertisement

Default settings
E.6 Dosing
HMI menu ID
Parameter
2.5.9.3
Amount
2.5.9.3
Amount
2.5.9.3
Amount
2.5.9.4
Decimal Places
2.5.9.5
Calibration
2.5.9.5.2
Fixed Compensation
2.5.9.5.2
Fixed Compensation
2.5.9.5.2
Fixed Compensation
2.5.9.6
Valve Control
2.5.9.6.1
Stage Setup Format
2.5.9.6.2
Stage 1 Primary Open
454
Description
Define the filling amount. The
valve will be closed and filling
stopped when defined amount is
reached.
Define the filling amount. The
valve will be closed and filling
stopped when defined amount is
reached.
Define the filling amount. The
valve will be closed and filling
stopped when defined amount is
reached.
Define the decimal point for the
process value displayed in the
Local User Interface.
Define the fixed pre-stop value.
The value is subtracted from
"Amount" to compensate for ap‐
plication delays in closing the
valves.
Define the fixed pre-stop value.
The value is subtracted from
"Amount" to compensate for ap‐
plication delays in closing the
valves.
Define the fixed pre-stop value.
The value is subtracted from
"Amount" to compensate for ap‐
plication delays in closing the
valves.
Specify if valve control parame‐
ters are defined as absolute or
relative values (% of dosing
amount).
Define the dosing amount value
to open the primary valve. The
value is in % of defined dosing
amount.
Default value
Value range
[unit]
1.0
[kg]
1.00
L
1.00
SL
2
● Auto
● No Decimal
place
● 1
● 2
● 3
● 4
● 5
● 6
[kg]
L
SL
Relative
● Relative
● Absolute
0
0.0 to 100.0
[%]
FC430 (From firmware 4.0)
Operating Instructions, 06/2017, A5E39789392-AA
Access
level
User
User
User
User
User
User
User
User
User