Siemens SITRANS FC430 Operating Instructions Manual page 200

Coriolis flowmeters
Hide thumbs Also See for SITRANS FC430:
Table of Contents

Advertisement

Diagnosing and troubleshooting
10.1 Device status symbols
NAMUR status signals
Icon
Priority
level
1
3
4
2
Info icons
Icon
Name
Read only
Data exchange
Custody transfer enabled
Memory card
Memory card
Memory card
Memory card
The table explains the meaning of possible device status and actions for the user or service.
Depending on the type and availability of local display of the device, status is shown as
SIMATIC and/or NAMUR NE107 symbol. Additionally the symbol for the according device
status is visualized in remote engineering, asset management or process control systems.
Refer to the manual of your automation system how diagnostics states are displayed.
200
Name
Description
Failure
● Failure
● Invalid output signal
Out of specification
● Out of specification
● Device will still work, but output signals may be invalid
Maintenance required
● Maintenance request
● Measured signal still valid
Function check
● Function check
● Output signal temporarily invalid
Description
● Write access disabled
● Indicate read only parameters
● Device is communicating
N/A
Capacity is 25 % used
Capacity is 50 % used
Capacity is 75 % used
Capacity is 100 % used
FC430 (From firmware 4.0)
Operating Instructions, 06/2017, A5E39789392-AA