Diagnosing and troubleshooting

10.3 Operation troubleshooting

ID/
Diagnostic
Icons
223
Modbus
285
Datalogging
286
Datalogging
287
Datalogging
288
Frame temperature simulated
298
Standard volume flow simulated
10.3
Operation troubleshooting
10.3.1
How do I copy application setup from one device to another?
1. Remove the SensorFlash from the source device and insert the SensorFlash into the
destination device. The destination device disables the backup and signals an alarm.
2. Enter menu item 3.3.1 Copy configuration, select "OK" and press
and move all the application setup parameters from the SensorFlash to the device. Backup
is still disabled and alarm signaled.
3. Remove the SensorFlash from the destination device and insert the original SensorFlash.
The device synchronizes the parameters to the SensorFlash and the alarm is cleared.
222
Action
Invalid coil configuration. Modbus coils are not
configured correctly. Check coil allocation.
DATALOGGING 30 DAYS LEFT
DATALOGGING 7 DAYS LEFT
DATALOGGING SDCARD FULL
Value simulated. Disable 'Simulation' to return
to normal operation.
Value simulated. Disable 'Simulation' to return
to normal operation.
Comment
to execute the copying
FC430 (From firmware 4.0)
Operating Instructions, 06/2017, A5E39789392-AA