Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
ID/
Diagnostic
Icons
40
Temperature measurement fault Turn off the power, wait 5 seconds and turn on
41
Temperature measurement fault Turn off the power, wait 5 seconds and turn on
42
Flow values not valid
43
Flow values not valid
44
Flow values not valid
45
Flow values not valid
46
Invalid calibration data
47
Invalid compensation data
49
Malfunction in Pickup Amplitude Contact your local Siemens representative
204
Action
the power again. If the failure continues then
contact your local Siemens representative
the power again. If the failure continues then
contact your local Siemens representative
Can be due to problems with measured fluid or
hardware malfunction. If the failure continues
then contact your local Siemens representative
Can be due to problems with measured fluid or
hardware malfunction. If the failure continues
then contact your local Siemens representative
Can be due to problems with measured fluid or
hardware malfunction. If the failure continues
then contact your local Siemens representative
Can be due to problems with measured fluid or
hardware malfunction. If the failure continues
then contact your local Siemens representative
Contact your local Siemens representative for
recalibration
Contact your local Siemens representative
Comments
FC430 (From firmware 4.0)
Operating Instructions, 06/2017, A5E39789392-AA