Push the plug into the »
in tightly.
Only alarm signals will be transferred through the nurse
call system.
Warning messages appear with 3 exclamation marks in the
upper line on screen; see page 31. Caution and advisory
messages will not be transferred. The nurse call system will
also be activated when the original alarm tone device is
defective, or the power supply to the device being used is cut
off.
Test the connected nurse call system for proper
functioning.
The installation of the nurse call system does not replace
the regular Savina screen surveillance.
Regularly check on-screen messages.
Technical data
Potential-free DC contact
Input voltage
Input current
Switching power

Checking readiness for operation

to be carried out after preparation.
The following functions are tested during this check:
— Operation of lamps/LEDs, displays and the audible alarm
— Ventilation function
— PEEP function
— Expiratory volume measurement MV
— Measurement of inspiratory O
— Lower alarm limit MV
— Lower alarm limit Paw
— Mains power failure alarm
— O
alarm
2
« socket at the back and screw
max. 40 V =
max. 500 mA
max. 15 W
concentration FiO
2
2
1
5
3
Preparation
93