Fitting the expiratory valve
1 Turn the locking ring (E) as far as possible to the
left.
2 Push the expiratory valve into the fitting.
3 Turn the locking ring (E) as far as it will go to the
right until it clicks audibly into place.
4 Check that it is properly secured by gently
pulling on the expiratory valve.
Additional information
Savina 300 can be equipped with the MP01781
expiratory filter.
Prerequisite: Savina 300 must be mounted on the
trolley.
For more information, see the instructions for use
for "Expiratory filter".
Instructions for use Savina 300 SW 5.n
E

Fitting the flow sensor

WARNING
Risk of fire
Residual vapors of highly flammable disinfec-
tants (e.g., alcohols) and deposits that were
not removed during reprocessing may ignite
when the flow sensor is in use.
– Ensure particle-free cleaning and disinfec-
tion.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor check for
visible damage and soiling, such as resid-
ual mucus, medication aerosols, and parti-
cles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
Prerequisite: The flap is open.
A
B
1 Push the socket (A) as far to the left as it will go.
2 Insert the flow sensor (B) into the socket with
the plug facing towards the device and push it
into the socket as far as it will go.
Assembly and preparation
53