Group
Syntax
Arguments
Examples
TRIGger:A:PULSEWidth:WHEn
Group
Syntax
Related Commands
Arguments
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual
Trigger
TRIGger:A:PULSEWidth:SOUrce {{CH1|CH2|CH3|CH4|
D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|D12|D13|D14|D15|LINE|EXT}
TRIGger:A:PULSEWidth:SOUrce?
specifies an analog input channel as the A edge trigger source.
CH1–CH4
specifies an digital input channel as the A edge trigger source.
D0–D15
specifies an external trigger using the Aux In connector located on the front
EXT
panel of the oscilloscope.
specifies AC line voltage.
LINE
TRIGGER:A:PULSEWIDTH:SOURCE CH1
TRIGGER:A:PULSEWIDTH:SOURCE?
indicating that channel 1 is the pulse width source.
SOURCE CH1
Sets or returns whether to trigger on a pulse that meets, falls outside, or within
the specified range of limits.
Trigger
TRIGger:A:PULSEWidth:WHEn {LESSthan|MOREthan|EQual|UNEQual}
TRIGger:A:PULSEWidth:WHEn?
TRIGger:A:PULSEWidth:WIDth
argument sets the oscilloscope to trigger if a pulse is detected with
LESSthan
width less than the time set by the
argument sets the oscilloscope to trigger if a pulse is detected with
MOREthan
width more than the time set by the
argument sets the oscilloscope to trigger if a pulse is detected with width
EQUal
equal to the time period specified in
a ±5% tolerance.
argument sets the oscilloscope to trigger if a pulse is detected with
UNEQual
width greater than or less than (but not equal) the time period specified in
TRIGger:A:PULSEWidth:WIDth
Commands Listed in Alphabetical Order
sets channel 1 as the pulse width source.
might return
:TRIGGER:A:PULSEWIDTH:
TRIGger:A:PULSEWidth:WIDth
TRIGger:A:PULSEWidth:WIDth
TRIGger:A:PULSEWidth:WIDth
within a ±5% tolerance.
command.
command.
within
2-329