Commands Listed in Alphabetical Order
Arguments
SEARCH:SEARCH<x>:TRIGger:A{:TRANsition|:RISEFall}:SOUrce
Group
Syntax
Arguments
SEARCH:SEARCH<x>:TRIGger:A{:TRANsition|:RISEFall}:WHEn
Group
Syntax
Arguments
2-282
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual
specifies that a pulse edge must traverse from the lower (most
POSitive
negative) to higher (most positive) level for transition triggering to occur.
specifies that a pulse edge must traverse from the upper (most positive)
NEGative
to lower (most negative) level for transition triggering to occur.
specifies either positive or negative polarity.
EITher
Sets or returns the source setting for an transition trigger search to determine
where to place a mark. <x> is the search number.
Search
SEARCH:SEARCH<x>:TRIGger:A{:TRANsition|:RISEFall}:SOUrce
{CH1|CH2|CH3|CH4|MATH}
SEARCH:SEARCH<x>:TRIGger:A{:TRANsition|:RISEFall}:SOUrce?
specifies one input channel as the edge source.
CH1–CH4
specifies the math waveform as the search source.
MATH
Sets or returns the condition setting for an transition trigger search to determine
where to place a mark. <x> is the search number.
Search
SEARCH:SEARCH<x>:TRIGger:A{:TRANsition|:RISEFall}:WHEn
{SLOWer|FASTer|EQual|UNEQual}
SEARCH:SEARCH<x>:TRIGger:A{:TRANsition|:RISEFall}:WHEn?
sets the trigger to occur when the transitioning signal is faster than the set
FASTer
volts/second rate.
sets the trigger to occur when the transitioning signal is slower than
SLOWer
the set volts/second rate.
sets the trigger to occur when the transitioning signal is equal to the set
EQual
volts/second rate within a ±5% tolerance.