Commands Listed in Alphabetical Order
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit
Group
Syntax
Arguments
Examples
TRIGger:A:LOGIc:PATtern:WHEn:MORELimit
Group
Syntax
Arguments
Examples
2-324
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual
Sets or returns the maximum time that the selected pattern may be true and still
generate an A logic pattern trigger.
Trigger
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit <NR3>
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit?
specifies the maximum amount of time to hold the pattern true.
<NR3>
TRIGGER:A:LOGIC:PATTERN:WHEN:LESSLIMIT 10.0E+00
maximum time that the selected pattern may hold true (and generate an A logic
pattern trigger) to 10 s.
TRIGGER:A:LOGIC:PATTERN:WHEN:LESSLIMIT?
TRIGGER:A:LOGIC:PATTERN:WHEN:LESSLIMIT 8.0000E-9
the selected pattern may hold true for up to 8 ns and still generate an A logic
pattern trigger.
Sets or returns the minimum time that the selected pattern may be true and still
generate an A logic pattern trigger.
Trigger
TRIGger:A:LOGIc:PATtern:WHEn:MORELimit <NR3>
TRIGger:A:LOGIc:PATtern:WHEn:MORELimit?
specifies the minimum amount of time to hold the pattern true.
<NR3>
TRIGGER:A:LOGIC:PATTERN:WHEN:MORELIMIT 10.0E+00
time that the selected pattern may hold true (and generate an A logic pattern
trigger) to 10 s.
TRIGGER:A:LOGIC:PATTERN:WHEN:MORELIMIT?
TRIGGER:A:LOGIC:PATTERN:WHEN:MORELIMIT 8.0000E-9
the selected pattern must hold true for at least 8 ns to generate an A logic pattern
trigger.
sets the
might return
indicating that
sets the minimum
might return
indicating that