To Install a Packed Column Adapter on a Detector Fitting
WA RN IN G
C AU TI O N
Maintaining Your GC
Gather the following:
1
•
7/16-inch, 9/16-inch, and 1/2-inch wrenches
•
Vespel/graphite ferrule (See
the Purged Packed Inlet"
•
Brass column nut
•
Lint-free gloves
•
Adapter.
Select the appropriate adapter from one of the parts lists
shown below:
•
Consumables and Parts for the FID
only be installed on an adaptable FID.)
•
Consumables and Parts for the TCD
•
Consumables and Parts for the NPD
•
Consumables and Parts for the FPD
•
For µECD, 1/4-inch packed columns require no adapter.
For 1/8-inch packed columns, see
for the
µECD.
Load the
GC maintenance method
2
become ready.
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Assemble a nut and a ferrule onto the adapter.
3
Maintaining the Purged Packed Inlet
"Consumables and Parts for
on page 60.)
(Packed columns can
+
Consumables and Parts
and wait for the GC to
5
83