To Replace The Fpd + Ignitor - Agilent Technologies 7820A Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

12
Maintaining the FPD
To Replace the FPD
WA RN IN G
246
+
+
Ignitor
Gather the following:
1
Ignitor replacement kit. (See
the FPD
Torx screwdrivers, T-20 and T-10
5/16-inch nut driver (or wrench)
Load the
2
become ready.
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
3
Remove the FPD
page 244.)
Loosen the collet screw holding the cable assembly to the
4
ignitor. Remove the collet and cable assembly.
5
Use a nut driver to loosen and remove the glow plug.
If using a 5/16-inch wrench, you may need to remove the
exit tube assembly using a T-10 Torx driver
+"
on page 234.)
GC maintenance method
+
cover. (See
"To Remove the FPD
"Consumables and Parts for
and wait for the GC to
+ Cover"
Maintaining Your GC
on