To Bakeout the NPD

WA RN IN G
WA RN IN G
Maintaining Your GC
If using hydrogen as a carrier gas, turn off the hydrogen supply
and cap the end of the column to prevent an oven explosion.
Bakeout the NPD with the column installed or uninstalled. If
1
uninstalled, gather the following (see
Parts for the NPD"
Capillary adapter (adaptable NPD only)
Column nut
No-hole ferrule
Load the
GC maintenance method
2
become ready.
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Set the bead voltage to 0, then turn it Off.
3
If the column is uninstalled, plug the detector connection
4
with the capillary adapter, column nut, and no-hole ferrule.
Maintain inert carrier gas flow through the column, or
remove the column from the GC.
Set normal operating flows.
5
Set the detector temperature at 25 °C above the typical
6
method setpoint temperature.
7
Set the oven temperature to 250 °C or 25 °C above the
normal maximum operating temperature. Do not exceed the
column's temperature limit.
Hold at temperature for 15 to 30 minutes.
8
9
If the column is not installed in the NPD, install it. (See
Install a Capillary Column in the NPD"
Restore the analytical method and allow the NPD to
10
equilibrate at operating temperatures and flows for 10 to
30 minutes.
Check the NPD leakage current. See
11
Leakage Current"
Start the NPD bead Auto Adjust process. See the
12
Advanced User Guide
"Consumables and
on page 208):
and wait for the GC to
"To Check the NPD
on page 230.
for details.
11
Maintaining the NPD
"To
on page 216.)
7820 Series
231