To Attach a Capillary Column Adapter on an Adaptable FID

WA RN IN G
C AU TI O N
Maintaining Your GC
Gather the following materials:
1
Adapter (See
"Consumables and Parts for the FID"
page 146.)
1/4-inch brass nut
1/4-inch Vespel/graphite ferrule
Column cutter
1/4-inch wrench
9/16-inch open-end wrench
Lint-free gloves
Load the
GC maintenance method
2
become ready.
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Assemble the nut and ferrule onto the adapter.
3
1/4-inch ferrule
1/4-inch nut
Adapter
Insert the adapter straight into the detector base as far as
4
possible.
Hold the adapter in this position and finger-tighten the nut.
5
Maintaining the FID
on
and wait for the GC to
8
153