Maintaining Your GC
Wearing gloves, inspect the O-ring and replace, if necessary.
5
(See
"To Change the O-Ring on the Packed Column Inlet"
page 101.)
6
Wearing gloves, grasp the flared end (top) of the replacement
glass liner with tweezers and install it in the inlet.
Install the top insert weldment and hand-tighten firmly.
7
Restore the analytical method.
8
Reset the EMF counter. See
9
Operation Manual.
Maintaining the Packed Column Inlet
To Reset an EMF Counter
6
on
in the
103