Maintaining the NPD

To Attach a Capillary Column Adapter on an Adaptable NPD

WA RN IN G
WA RN IN G
C AU TI O N
214
Gather the following materials:
1
Adapter (See
"Consumables and Parts for the NPD"
page 208.)
1/4-inch nut
1/4-inch ferrule
Column cutter
1/4-inch wrench
9/16-inch open-end wrench
Lint-free gloves
Load the
GC maintenance method
2
become ready.
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Assemble a brass nut and Vespel/graphite ferrule onto the
3
adapter.
1/4-inch ferrule
1/4-inch nut
Adapter
on
and wait for the GC to
Maintaining Your GC