Maintaining Your GC
Install the collector nut over the castle and tighten firmly.
d
The seal should be airtight. Maintain the orientation of
the ignitor hole with the base as shown below.
16
Install the FID ignitor.
a
Insert the ignitor and copper seal into the threaded hole
of the castle. Keep the mating threads clean.
Tighten the ignitor with a wrench. Ignition requires a
b
good electrical contact that is free of any dirt.
8
Maintaining the FID