To Install the FID Insulation Cup Assembly (Adaptable FID Only)

WA RN IN G
Maintaining Your GC
Gather the following:
1
Insulation (See
page 146.)
Insulation cup assembly
Load the
GC maintenance method
2
become ready.
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Assemble the insulation in the cup. Line up the slots in the
3
insulation with the slot in the cup.
Capillary columns should be attached to the detector before
installing the cup. When attaching a packed column to the
detector, cap the detector fitting before installing the cup to
prevent insulation contamination of the detector.
4
Push the wire spring lever to the right to uncover the hole.
"Consumables and Parts for the FID"
and wait for the GC to
8
Maintaining the FID