Selecting an FID Jet

Table 24
Jets for capillary optimized fittings
Figure 1 ID
Jet type
1
Capillary
2
High-temperature
(use with simulated
distillation)
Maintaining Your GC
Open the oven door and locate the column connection fitting at
the base of the detector. It will look like either a capillary
optimized fitting or an adaptable fitting.
Capillary optimized fitting
• If you have an application that tends to clog the jet, select a
jet with a wider tip id.
• When using packed columns in high column-bleed
applications, the jet tends to clog with silicon dioxide.
• In simulated distillation applications, the high-boiling
hydrocarbons tend to clog the jet.
For capillary optimized fittings, select a jet from
adaptable fittings, select a jet from
Part number
G1531-80560
G1531-80620
Figure 2
Capillary optimized jets
Maintaining the FID
Adaptable fitting
Detector fitting
Adapter
Table
25.
Jet tip id
Length
0.29 mm (0.011 inch)
43 mm
0.47 mm (0.018 inch)
48 mm
8
Table
24. For