Agilent Technologies 7820A Manual page 159

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

C AU TI O N
Maintaining Your GC
This step exposes the interconnect spring. Be careful not to touch or
disfigure the spring while working on the FID. Any dirt or bending
will reduce the sensitivity of your detector.
5
Lift and remove the assembly from the pallet.
Remove the ignitor cable assembly from the new collector
6
assembly, if present.
Remove any protective caps from the new collector assembly,
7
if present.
8
Lower the new collector assembly into the housing.
Insert the three screws and tighten (to 18 inch-pounds).
9
Connect the ignitor extension cable.
10
Verify assembly:
11
Check the FID leakage current. (See
a
Leakage Current"
b
Check the FID baseline. (See
on page 174.)
Maintaining the FID
"To Check the FID
on page 173.)
"To Check the FID Baseline"
8
159