To Install an Adapter on the Packed Column Inlet

WA RN IN G
C AU TI O N
Maintaining Your GC
Gather the following:
1
Brass tubing nut (see
Packed Column Inlet"
Adapter (1/8-inch packed or 1/4-inch packed)
7/16-inch and 9/16-inch wrench
Vespel/graphite ferrule
Methanol
Lint-free gloves
Load the
inlet maintenance method
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
3
Clean the end of the adapter with a lint-free cloth and
methanol to remove contamination such as fingerprints.
Place the tubing nut and Vespel/graphite ferrule on the
4
adapter.
Maintaining the Packed Column Inlet
"Consumables and Parts for the
on page 92.)
and wait for the GC to