Maintaining Your GC
Fill an ultrasonic cleansing bath with aqueous detergent and
6
place the spring and insert into it. Sonicate for 1 minute.
7
Drain the aqueous detergent and fill the bath with distilled
water. Sonicate for 1 minute.
Remove the spring and insert from the bath and rinse
8
thoroughly with water and methanol.
9
Dry the spring and insert with compressed air or nitrogen.
10
Install the insert. (See
on page 132.)
Install the column. (See
11
the COC Inlet"
on page 125.)
Maintaining the COC Inlet
"To Install an Insert on the COC Inlet"
"To Install a Capillary Column with
7
135