Maintaining Your GC
Remove the existing insert from the inlet by gently pushing it
6
out from below with a wire or piece of column. Store the
insert for possible later use.
Check that the insert is the correct size for both the needle
7
and column. (See
"To Check the Needle-to-Column Size on
the COC Inlet"
on page 128.)
Lower the new insert straight into the inlet from the top. The
8
insert can be installed either end up.
9
Install the spring on top of the insert.
10
Install the septum and septum nut or duckbill septum and
cooling tower assembly and finger-tighten.
Install the column. (See
11
the COC Inlet"
on page 125.)
Maintaining the COC Inlet
"To Install a Capillary Column with
7
133