To Bakeout The Μecd - Agilent Technologies 7820A Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

To Bakeout the µECD
WA RN IN G
C AU TI O N
WA RN IN G
WA RN IN G
Maintaining Your GC
Detector disassembly and/or cleaning procedures other than
thermal should be performed only by personnel trained and
licensed appropriately to handle radioactive materials. Trace
amounts of radioactive
procedures, causing possible hazardous exposure to b- and
x-radiation.
To prevent possible hazardous contamination of the area with
radioactive material, the detector exhaust vent always must be
connected to a fume hood or otherwise vented in compliance with
the latest revision of 10 CFR Part 20, or with state regulations with
which the Nuclear Regulatory Commission has entered into an
agreement (USA only). For other countries, consult with the
appropriate agency for equivalent requirements.
Gather the following:
1
Column nut and no-hole ferrule (See
Parts for the µECD"
Blanking nut with any column ferrule
2
With the detector and oven at normal operating
temperatures, show the detector output. Using the software
keyboard, press [Front Det] or [Back Det], then scroll to Output.
Alternately, use the arrows keys on the GC to scroll Output.
Note the value of Output for later comparison.
Load the
GC maintenance method
3
become ready.
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Maintaining the µECD
63
Ni may be removed during other
"Consumables and
on page 194.)
and wait for the GC to
10
205