7
Maintaining the COC Inlet
To Change a Septum on the COC Inlet
WA RN IN G
C AU TI O N
130
Gather the following:
1
•
Replacement septum. (See
the COC Inlet"
on page 120.)
•
Tweezers
•
A thin wire (0.2-inch diameter) for removing septum from
inlet
•
Lint-free gloves
Load the
inlet maintenance method
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Replace the septum.
3
•
If you are using a septum nut, grasp the knurling and
unscrew. Remove the old septum with tweezers. Use
tweezers to install a new septum. Push the septum into
the septum nut until properly seated. Firmly tighten the
nut.
For 250/320-µm
For 530-µm automated
automated injections
injections
"Consumables and Parts for
and wait for the GC to
Maintaining Your GC