To Install The Optional Tcd Capillary Column Adapter - Agilent Technologies 7820A Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

9
Maintaining the TCD

To Install the Optional TCD Capillary Column Adapter

C AU TI O N
186
Gather the following:
1
Capillary column adapter (See
for the TCD"
on page 182.)
1/4-inch and 7/16-inch wrenches
1/8-inch brass nut
1/8-inch Vespel ferrule
Lint-free gloves
Load the
GC maintenance method
2
become ready.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
3
Assemble the brass nut and ferrule onto the capillary column
adapter.
Ferrule, 1/8-inch
Nut, 1/8-inch
Capillary column adapter
4
Install the adapter assembly into the detector fitting and
tighten finger-tight. Use a wrench to tighten until snug.
"Consumables and Parts
and wait for the GC to
Maintaining Your GC