Maintaining Your GC
Use compressed air or nitrogen to blow away the pieces of
7
steel wool and septum.
Replace the insert retainer nut, tightening it to firm
8
finger-tightness. Do not overtighten.
9
Firmly press the new septum or Merlin Microseal into the
fitting. (See
"To Change the Septum on the Split/Splitless
Inlet"
on page 44.)
10
Replace the septum retainer nut or Merlin cap and
finger-tighten. (See
Split/Splitless Inlet"
11
Restore the analytical method.
Maintaining the Split/Splitless Inlet
"To Change the Septum on the
on page 44.)