Maintaining Your GC
Score the column using a glass scribing tool. The score must
4
be square to ensure a clean break.
Break off the column end by supporting it against the column
5
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain that there are no burrs or jagged
edges.
Wipe the column walls with a tissue dampened with
6
isopropanol to remove fingerprints and dust.
7
Insert the column into the detector until it bottoms.
8
Slide the column nut and ferrules up the column to the
detector and finger-tighten the nut.
Pull out 1 mm of column. Tighten the nut an additional
9
1/4 turn with a wrench or until the column does not move.
Maintaining the TCD
Bad
Good
9
185