To Install a Capillary Column with the COC Inlet

WA RN IN G
WA RN IN G
C AU TI O N
Maintaining Your GC
Gather the following:
1
Column nut and ferrule. (See
the COC Inlet"
on page 120.)
Column cutter
1/4-inch and 5/16-inch wrenches
Lint-free gloves
Load the
inlet maintenance method
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
3
Before installing the column, be sure the correct insert is
installed for the needle and column. (See
on the COC Inlet"
4
Place a capillary column nut and ferrule on the column.
Ferrule
Column nut
5
Score the column using a glass scribing tool. The score must
be square to ensure a clean break.
Maintaining the COC Inlet
"Consumables and Parts for
and wait for the GC to
on page 132.)
7
"To Install an Insert
125