To Attach a Capillary Column to the FPD

WA RN IN G
WA RN IN G
C AU TI O N
Maintaining Your GC
Gather the following:
1
Column measuring tool (See
the FPD"
on page 252.)
Column cutter
1/4-inch and 7/16-inch wrenches
Column nut
Ferrule
Capillary column
Lint-free gloves
Load the
GC maintenance method
2
become ready.
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Assemble a septum, column nut, and ferrule on the end of
3
the column.
Insert the end of the column through the column measuring
4
tool so that the end protrudes beyond the tool.
Maintaining the G3435A/G3436A FPD
"Consumables and Parts for
and wait for the GC to
13
257