Maintaining the G3435A/G3436A FPD

To Remove the FPD Vent Tube

C AU TI O N
WA RN IN G
262
Gather the following:
1
T-20 Torx screwdriver
9/16-inch wrench
Load the
GC maintenance method
2
become ready.
When turning off the GC, first turn off the flame to prevent
condensation from dripping into the jet and column.
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
3
Drain any water from the flexible vent tubing and discard.
4
Open the FPD cover.
Remove the flexible tubing from the FPD vent tube.
5
Loosen and remove the vent tube assembly with a wrench.
6
Remove the screws securing the FPD cover.
7
The single-wavelength detector has two screws at the
bottom of the left side (top photo below) and two screws
at the top of the right side (bottom photo below).
and wait for the GC to
Maintaining Your GC