To Condition A Capillary Column - Agilent Technologies 7820A Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

3
Maintaining Capillary Columns

To Condition a Capillary Column

WA RN IN G
WA RN IN G
WA RN IN G
28
Gather the following:
1
One 7/16-inch, and 1/4-inch wrenches
No-hole ferrule (See
Columns"
on page 24.)
Column nut
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
Load the
GC maintenance method
2
become ready.
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Install the column into the inlet using the new ferrules. See:
3
To Install a Capillary Column with the Split/Splitless Inlet
To Install a Capillary Column with the Purged Packed
Inlet
To Install a Capillary Column with the COC Inlet
"Consumables and Parts for
and wait for the GC to
Maintaining Your GC