Agilent Technologies 7820A Manual page 261

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

C AU TI O N
C AU TI O N
Maintaining Your GC
Sulfur filter
Filter spacer
Do not use cleaning fluids. Cleaning fluids will damage lens
coatings.
7
Clean the new filter with lens tissue.
Filters are designed for the light of the flame to pass through in a
specific direction. The triangle (on the edge of the phosphorus filter)
and the arrow (on the edge of the sulfur filter) should face away
from the flame and toward the PMT.
Install the filter in the filter housing. Install the sulfur filter
8
spacer, if necessary.
9
Replace the PMT assembly and secure with the spring.
Restore the analytical method.
10
Maintaining the G3435A/G3436A FPD
13
261

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents