Agilent Technologies 7820A Manual page 66

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

5
Maintaining the Purged Packed Inlet
66
1%2 mm
Thread the column nut into the inlet adapter but do not
10
tighten.
11
Adjust the column position so that the septum is even with
the bottom of the column nut. Finger-tighten the column nut
until it begins to grip the column.
12
Tighten the column nut an additional 1/4 to 1/2 turn with a
wrench so that the column cannot be pulled from the fitting
with gentle pressure.
Configure the new column.
13
14
Condition the column per the manufacturer's
recommendation. (See
Install the column into the detector. See:
15
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the µECD
To Attach a Capillary Column to the FPD
To Condition a Capillary
+
Column.)
Maintaining Your GC