To Prepare the Column Ends for a CFT Fitting

Materials needed
• SilTite ferrules:
• Swaging nut
• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Procedure
1
2
Figure 5
The SilTite ferrules are delicate. Follow the instructions in the next step very carefully
CA UT IO N
to avoid overtightening.
5977 Series MSD Operation Manual
0.1- to 0.25-mm columns, pkg of 10 (5188-5361)
0.32-mm columns, pkg of 10 (5188-5362)
0.53-mm columns, pkg of 10 (5188-5363)
Obtain the appropriate metal ferrule.
Pass the tubing end through the internal nut and SilTite ferrule leaving
approximately 1 cm of fused silica tubing protruding beyond the ferrule.
Thread the ferrule pre-swaging tool onto the internal nut with the tube
protruding (see
Figure
Assemble the internal nut, ferrule, and swaging nut
5).
Installing GC Columns
2
41