Maintaining the Purged Packed Inlet
76
Wearing gloves, inspect the O-ring and replace, if necessary.
5
(See
"To Change the O-Ring on the Purged Packed Inlet"
page 74.)
6
Wearing gloves, grasp the flared end (top) of the replacement
glass liner with tweezers and install it in the inlet. If the glass
liner does not seat properly because a capillary column is
installed, remove the column, install the glass liner, and
replace the column. (See
the Purged Packed Inlet"
Install the top insert weldment and hand-tighten firmly.
7
Restore the analytical method.
8
Reset the EMF counter. See
9
Operation Manual.
"To Install a Capillary Column with
on page 64.)
To Reset an EMF Counter
Maintaining Your GC
on
in the