To Change The Glass Liner On The Purged Packed Inlet - Agilent Technologies 7820A Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

To Change the Glass Liner on the Purged Packed Inlet

WA RN IN G
Top insert weldment
Flared end
Glass liner
C AU TI O N
Maintaining Your GC
Gather the following:
1
Replacement glass liner (See
the Purged Packed Inlet"
9/16-inch wrench
Lint-free gloves
Load the
inlet maintenance method
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
3
Loosen the top insert weldment to remove the top portion of
the inlet.
Use a thin wire or wood splint to carefully lift and remove
4
the old glass liner.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Purged Packed Inlet
"Consumables and Parts for
on page 60.)
and wait for the GC to
5
75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents