To Perform Maintenance on the FID Collector Assembly

NO T E
C AU TI O N
WA RN IN G
Maintaining Your GC
Perform only the steps and gather only the parts that apply to the desired
maintenance task(s).
Gather the following:
1
Replacement ignitor assembly (See
Parts for the FID"
Replacement ignitor castle
Two collector insulators
Collector
Spring washer
Gasket
T-20 Torx screwdriver
1/4-inch nut driver
Tweezers
5/16-inch wrench
Lint-free gloves
Clean cloth
To avoid contaminating the FID, wear clean, lint-free gloves when
handling the collector assembly.
Load the
GC maintenance method
2
become ready.
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
Maintaining the FID
"Consumables and
on page 146.)
and wait for the GC to