Maintaining Your GC
Install the adapter, if used. (See
12
Purged Packed Inlet"
Install the glass liner and O-ring. (See
13
Liner on the Purged Packed Inlet"
14
Install the top insert weldment and finger-tighten.
Install the septum and septum nut. (See
15
Septum on the Purged Packed Inlet"
Attach the column. (See
16
the Purged Packed Inlet"
17
Check for leaks.
Restore the analytical method.
18
Set the septum and glass liner counters.
19
Maintaining the Purged Packed Inlet
"To Install an Adapter on the
on page 72.)
"To Change the Glass
on page 75.)
"To Change the
on page 68.)
"To Install a Capillary Column with
on page 64.)