Agilent Technologies 7820A Manual page 217

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

C AU TI O N
Maintaining Your GC
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Place a septum, capillary column nut, and ferrule on the
3
column.
Ferrule
Column nut
Septum
4
Score the column using a glass scribing tool. The score must
be square to ensure a clean break.
Break off the column end by supporting it against the column
5
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
6
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
Install the capillary column.
7
Maintaining the NPD
Bad
Good
11
217