To Install A Packed Column Adapter In The Fpd - Agilent Technologies 7820A Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

To Install a Packed Column Adapter in the FPD

NO T E
FPD+ packed column adapter view
Installation
WA RN IN G
C AU TI O N
Maintaining Your GC
If you do not use a Packed column, ignore this section.
This procedure applies to installing the 1/8-inch packed metal
column adapter into the FPD
Gather the following:
7/16-inch wrench
5/16-inch wrench
1/8-inch nut and ferrule for a 1/8-inch packed column
Lint-free gloves
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the FPD
+
5062-3538 Adapter ferrule (10/pk)
G3435-20375 Adapter nut
G3435-60340 1/8-inch Packed column adapter
+
column fitting.
+
12
237