To Install a Capillary Column Adapter in the FPD

WA RN IN G
WA RN IN G
C AU TI O N
Maintaining Your GC
Gather the following:
1
FPD capillary column adapter (See
Parts for the FPD"
Column cutter
1/4-inch wrench
9/16-inch wrench
Metric ruler
1/8-inch nut
Lint-free gloves
Load the
GC maintenance method
2
become ready.
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Insert the capillary adapter into the 1/8-inch nut as shown,
3
then thread the nut onto the detector fitting.
Maintaining the G3435A/G3436A FPD
"Consumables and
on page 252.)
and wait for the GC to
13
255