Maintaining the Packed Column Inlet

To Install a Packed Column Adapter on a Detector Fitting

WA RN IN G
C AU TI O N
110
Gather the following:
1
7/16-inch, 9/16-inch, and 1/2-inch wrenches
Vespel/graphite ferrule (See
the Packed Column Inlet"
Brass column nut
Lint-free gloves
Adapter.
Select the appropriate adapter from one of the parts lists
shown below:
Consumables and Parts for the FID
only be installed on an adaptable FID.)
Consumables and Parts for the TCD
Load the
GC maintenance method
2
become ready.
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Assemble a nut and a ferrule onto the adapter.
3
Insert the adapter straight into the detector base as far as
4
possible.
"Consumables and Parts for
on page 92.)
(Packed columns can
and wait for the GC to
Maintaining Your GC