Maintaining the COC Inlet
126
Break off the column end by supporting it against the column
6
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
7
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
Gently insert the column into the inlet until it bottoms. You
8
should feel spring tension as you push up on the column. (Do
not withdraw the column.)
Insert the column nut into the inlet fitting and finger-tighten.
9
Bad
Good