To Install A Capillary Column With The Purged Packed Inlet - Agilent Technologies 7820A Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

5
Maintaining the Purged Packed Inlet

To Install a Capillary Column with the Purged Packed Inlet

WA RN IN G
WA RN IN G
C AU TI O N
64
Gather the following:
1
Column
Ferrule (See
"Consumables and Parts for the Purged
Packed Inlet"
on page 60.)
Column nut
Glass liner
Viton O-ring
0.53-mm column adapter
Septum
1/4-inch wrench
Metric ruler
Lint-free gloves
Load the
inlet maintenance method
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Install a 0.53-mm column adapter. (See
3
Adapter on the Purged Packed Inlet"
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Install a new Viton O-ring. (See
4
Purged Packed Inlet"
Place a septum, capillary column nut, and ferrule on the
5
column.
and wait for the GC to
"To Install an
on page 72.)
"To Change the O-Ring on the
on page 74.)