Maintaining the Purged Packed Inlet

To Install a Packed Column Adapter on a Detector Fitting

WA R N I N G
C A U T I O N
92
1
Gather the following:
7/16- inch, 9/16- inch, and 1/2- inch wrenches
Vespel/graphite ferrule (See
for the Purged Packed Inlet"
Brass column nut
Lint- free gloves
Adapter.
Select the appropriate adapter from one of the parts lists
shown below:
Consumables and Parts for the FID
can only be installed on an adaptable FID.)
Consumables and Parts for the TCD
Consumables and Parts for the NPD
Consumables and Parts for the FPD
For uECD, 1/4- inch packed columns require no
adapter. For 1/8- inch packed columns, see
Consumables and Parts for the
Load the
GC maintenance method
2
become ready.
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
3
Assemble a nut and a ferrule onto the adapter.
"Consumables and Parts
on page 68.)
(Packed columns
uECD.
and wait for the GC to