13
Maintaining the G3435A/G3436A FPD

To Install the FPD Vent Tube and Cover

266
Gather the following:
1
T-20 Torx screwdriver
9/16-inch wrench
Install the cover.
2
Single-wavelength detector:
a
Start the two screws on the right side of the cover.
b
Start and tighten the screws at the base on the left side.
c
Tighten the screws on the right side.
3
Install the vent tube assembly.
Reconnect the flexible tubing to the vent tube assembly and
4
route the open end to waste.
Close the FPD cover.