188
Break off the column end by supporting it against the column
5
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
6
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
Gently insert the column into the detector until it bottoms. Do
7
not attempt to force it further.
8
Slide the column nut and ferrule up the column to the
adapter and tighten the nut finger tight.
Bad
Good
Maintaining Your GC