Maintaining the TCD

To Install a Capillary Column in the TCD

WA RN IN G
WA RN IN G
C AU TI O N
184
Gather the following:
1
Front ferrule (See
on page 182.)
Back ferrule
Column nut
Column cutter
7/16-inch wrench
Lab tissue
Lint-free gloves
Load the
GC maintenance method
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Assemble the ferrules and 1/8-inch brass Swagelok nut on
3
the column.
Front ferrule
Back ferrule
Nut
"Consumables and Parts for the TCD"
and wait for the GC to
Maintaining Your GC