15
Parameter
Reference mark
Analog filter frequency
Terminating resistor
Error monitor
Counting direction
342
Explanation
Definition of the type of reference mark
Settings:
None: There is no reference mark
One: The encoder has one reference mark
Coded: The encoder has distance-coded reference marks
Default value: One
Frequency value of the analog low-pass filter (not with TTL)
Settings:
33 kHz: Suppression of interference frequencies above
33 kHz
400 kHz: Suppression of interference frequencies above
400 kHz
Default value: 33 kHz
Dummy load to avoid reflections
Settings: ON or OFF
Default value: ON
Monitoring of signal errors
Settings:
Off: Error monitoring not active
Amplitude: Error monitoring of signal amplitude
Frequency: Error monitoring of signal frequency
Frequency & amplitude: Error monitoring of signal amplitude
and signal frequency
Default value: Frequency & amplitude
Signal detection during axis movement
Settings:
Positive: The direction of traverse corresponds to the
counting direction of the encoder
Negative: The direction of traverse does not correspond to
the counting direction of the encoder
Default value: Positive