Determining the pixel sizes
When measuring with a VED sensor, the measurement is performed in the live
image on the product. To ensure that the size of the live image matches that of the
measured object, the pixel size must be determined for each magnification.
Further information:
150
Tap Settings in the main menu
Tap Sensors
Open in the sequence
Video edge detection (VED)
Pixel sizes
Tap Magnification
Select the desired magnification
In Calibration standard diameter, enter the diameter of
the desired circle, which is specified in the calibration chart
supplied with the calibration standard
Confirm entry with RET
Tap Start
The teach sequence is started and the wizard is displayed in
the Measure menu
Follow the instructions of the wizard
Tap Confirm to confirm that you have completed the wizard's
instructions
The teach sequence is complete
Tap Undo to repeat the teach sequence
Tap Close to close the wizard
Repeat the procedure to determine the pixel sizes for all
available magnifications
"Pixel sizes", page 329